Measuring colorimeter precision and accuracy
Jo Long — Tue, 2012-05-15 16:40
MethodsTo calculate the accuracy of the colorimeter we have used the ammonia colorimetric assay. A freshly opened 12.2 ppm ammonia standard solution (Ricca Chemical Company from Scientific Stategies Cat # 5450-4RC) was carefully diluted with distilled water as follows:
Each dilution was carried out in duplicate and the API ammonia test kit was used to develop 1 mL of each solution in triplicate. Absorbance was measured with the colorimeter red LED and data is shown in the Table below. |
|||||||||
Results
|
Sample Data
| Ammonia (ppm) | Absorbance #1 | Absorbance #2 | Absorbance #3 | Absorbance Mean | SD | *Colorimeter Reading | Accuracy (ppm) |
|---|---|---|---|---|---|---|---|
| 0.61 | 0.20 | 0.18 | 0.19 | 0.19 | 0.010 | 0.60 | -0.021 |
| 0.61 | 0.19 | 0.20 | 0.21 | 0.20 | 0.010 | 0.63 | 0.031 |
| 1.22 | 0.39 | 0.36 | 0.39 | 0.38 | 0.017 | 1.19 | -0.021 |
| 1.22 | 0.40 | 0.38 | 0.41 | 0.40 | 0.015 | 1.25 | 0.022 |
| 2.44 | 0.85 | 0.82 | 0.80 | 0.82 | 0.025 | 2.59 | 0.061 |
| 2.44 | 0.81 | 0.81 | 0.81 | 0.81 | 0.000 | 2.55 | 0.044 |